MEDIUM · CVSS 6.8 Published: 2026-06-02
CVE-2025-15653 — Technical Security Advisory & CVSS Profile
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Technical Metadata
| Identifier | CVE-2025-15653 |
|---|---|
| Base Severity / Score | MEDIUM (6.8 / 10.0) |
| CVSS Vector | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| Weakness Classification | CWE-668 |
| Archive Categories | MEDIUM Advisories · 2026 Index |
Operational Mitigation Guidance
Schedule vendor updates during routine patching windows and verify whether affected code paths are reachable in your deployment configuration. For architectural isolation principles, review per-process namespace isolation.
Upstream References
- https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf
- https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation
Related Security Advisories
CVE-2025-15642 MEDIUM 0
CVE-2025-1566 HIGH 7.5
CVE-2025-15641 MEDIUM 0
CVE-2025-15661 MEDIUM 6.5
CVE-2025-1571 MEDIUM 6.4
CVE-2025-1572 MEDIUM 6.5
CVE-2025-1576 MEDIUM 6.3